Validation of a Deep-Learning Coregistration Framework for Long-Axial-Field-of-View PET/CT Using Low-Radiation-Exposure Protocols Across Various Tracers
Clinical Snapshot
PICO Framework
| P — Population | Patients undergoing PET/CT imaging with long-axial-field-of-view systems using various tracers (89Zr-trastuzumab, 15O-H2O, 18F-MC225, and 18F-FDG) |
| I — Intervention | Deep-learning coregistration framework applied to low-dose PET and ultra-low-dose CT protocols |
| C — Comparator | Standard-dose PET and low-dose CT coregistration |
| O — Outcomes | Image alignment accuracy measured by Dice similarity coefficient, Jaccard similarity coefficient, Hausdorff distance, average surface distance, and SUV quantification accuracy |
Bottom Line
This validation study demonstrates that a deep-learning coregistration framework maintains robust performance for LAFOV PET/CT alignment even when using significantly reduced radiation protocols. The framework showed consistent accuracy across four different tracers and maintained quantitative precision with PET count reductions down to 12.5% of original levels and ultra-low-dose CT. While the study provides promising evidence for radiation dose optimisation in PET/CT imaging, the lack of explicit blinding and limited statistical precision reporting somewhat limit confidence in the findings. The technology appears ready for clinical implementation in centres with appropriate computational resources, potentially enabling safer imaging protocols particularly beneficial for research applications and repeat imaging scenarios.
Key Findings
P Value: Not reported in abstract
Effect Size: Consistent similarity metrics when comparing low-count reconstructed CT with reference CT
Primary Outcome: Robust coregistration performance across reduced PET counts and ultra-low-dose CT conditions
Nnt Or Sensitivity: Maintained quantitative accuracy across varying dose conditions as measured by SUV consistency
Confidence Interval: Not reported in abstract
Clinical Application
Requires LAFOV PET/CT systems and deep-learning computational infrastructure Relevant for Australian nuclear medicine departments with LAFOV systems. Supports ACPSEM radiation protection principles and could facilitate TGA-approved low-dose protocols Patients requiring PET/CT imaging where radiation dose reduction is prioritised, particularly in research settings and vulnerable populations
Abstract
The advent of long-axial-field-of-view (LAFOV) PET/CT systems has significantly improved whole-body imaging by providing higher sensitivity and extended torso coverage. However, PET/CT potential misalignment remains a challenge, potentially introducing artifacts and quantification errors. Moreover, PET protocols with reduced scanning duration and dose, as well as the use of ultra-low-dose CT (ULD-CT), are increasingly relevant in clinical practice and research. This study aimed to evaluate the robustness and generalizability of a deep-learning coregistration framework for PET/CT alignment using low-dose PET and ULD-CT across various tracers with a LAFOV system. Methods: In total, 63 scans with 4 different tracers (89Zr-trastuzumab, 15O-H2O, 18F-MC225, and 18F-FDG) were included to assess PET and CT alignment improvements. Further evaluation was performed by comparing CT scans coregistered to the original PET images (rCT) with those coregistered to low-count (50%, 25%, and 12.5% of original counts) PET images (LC-rCT). Dice similarity coefficient, Jaccard similarity coefficient, Hausdorff distance, and average surface distance were used as evaluation metrics. Furthermore, ULD-CT coregistered to PET (rULD-CT) was compared with low-dose CT coregistered to the same PET (rLD-CT) using the same metrics. PET accuracy was evaluated by calculating SUVs. Results: The robustness of the deep-learning coregistration framework was demonstrated in reduced PET counts and ULD-CT scenarios across tracers. All metrics indicated robust performance when comparing LC-rCT with rCT and rULD-CT with rLD-CT. Consistent SUVs across varying PET counts and ULD-CT conditions further validated the quantitative accuracy of the approach. Conclusion: This work highlights that neither the use of low-dose PET protocols nor ULD-CT compromise the performance of this coregistration framework across 4 tracers. For LAFOV PET, these findings support the feasibility of such a framework in imaging protocols with reduced radiation exposure.
References
- 1.Li, Z., Providência, L., Mohr, P., Mostafapour, S., Roya, M., Martinez-Lucio, T. S., Somsen, J. F., van Leeuwen, G. L., Salvi De Souza, G., van Sluis, J., Glaudemans, A. W. J. M., Dierckx, R. A. J. O., de Vries, J.-P. P. M., Luurtsema, G., Slart, R. H. J. A., Brouwers, A. H., Schleyer, P., Conti, M., Lammertsma, A. A., Schaefferkoetter, J., & Tsoumpas, C. (2026). Validation of a deep-learning coregistration framework for long-axial-field-of-view PET/CT using low-radiation-exposure protocols across various tracers. *Journal of Nuclear Medicine*, *67*(5). https://doi.org/10.2967/jnumed.125.270420
Related Research
European journal of radiology
A CT-based deep learning model to differentiate between benign and malignant adrenal lesions
3 Aug 2026
European respiratory review : an official journal of the European Respiratory Society
Prediction of PD-L1 expression in nonsmall cell lung cancer using artificial intelligence models based on radiomics: a systematic review and meta-analysis
10 July 2026
Radiation protection dosimetry
Impact of neutron sieve therapy for deep site in boron neutron capture therapy using Monte Carlo calculation
8 July 2026
This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service