Machine Learning-Enabled Gas Sensor Based on MOF-Derived In2O3-CuO for Exhaled CO Detection

ACS sensorsZhao, Fan, Jiang, Zhiyuan, Jiang, Xiangrui et al.26 June 2026DOI

Clinical Snapshot

35CEBM
Evidence: Weakother

PICO Framework

P — PopulationNeonates at risk of jaundice (hyperbilirubinaemia), with implied extension to clinical breath-analysis settings requiring CO biomarker detection
I — InterventionAn electronic nose (e-nose) system comprising five MOF-derived In2O3-CuO metal-oxide semiconductor sensors assembled into a microsensor array, augmented by a linear discriminant analysis (LDA) machine-learning algorithm with thermodynamic feature engineering
C — ComparatorNo direct clinical comparator; analytical performance benchmarked against single-sensor operation and against detection of the principal interferent gas acetone (C3H6O) in simulated breath matrices
O — OutcomesPrimary: qualitative discrimination of CO from acetone (cross-validation accuracy); Secondary: limit of detection (LOD), response/recovery times, reproducibility, moisture resistance, and stability of the 3In2O3-CuO sensor

Bottom Line

This proof-of-concept engineering study describes the fabrication and preliminary analytical characterisation of a five-sensor In2O3-CuO metal-oxide array, augmented by linear discriminant analysis, for detecting exhaled carbon monoxide as a biomarker of neonatal haemolysis and jaundice. The sensor achieves a limit of detection of 1 ppm CO with rapid response kinetics and demonstrates 82% cross-validation accuracy in discriminating CO from acetone in synthetic gas mixtures. While the materials science is technically sound and the clinical motivation is compelling, the study represents an early-stage laboratory proof of concept with no human data, no testing against real exhaled breath, and no comparison to established clinical standards. The 82% classification accuracy is insufficient for a neonatal screening application, and the absence of confidence intervals, formal statistical testing, and evaluation against the full spectrum of breath interferents are significant limitations. Senior clinicians should regard this as a promising but distant translational prospect. Substantial further work — including real breath matrix testing, clinical validation against serum bilirubin, and TGA regulatory evaluation — is required before this technology could be considered for Australian neonatal practice.

Evidence: Weak

Key Findings

  • P Value: Not reported

  • Effect Size: Not formally reported as an effect size; best-performing single sensor (3In2O3-CuO) demonstrated highest response magnitude to CO relative to acetone, with response defined as Ra/Rg (air resistance / gas resistance ratio)

  • Primary Outcome: Qualitative discrimination between CO and acetone (C3H6O) using a five-sensor In2O3-CuO array with LDA machine learning; cross-validation accuracy of 82% for CO identification across 1–50 ppm range

  • Nnt Or Sensitivity: Clinical sensitivity and specificity not calculable from this study; analytical LOD = 1 ppm CO; response time = 5 seconds; recovery time = 21 seconds; cross-validation classification accuracy = 82%

  • Confidence Interval: Not reported for any outcome measure

Clinical Application

Clinical feasibility is premature to assess. Significant engineering, regulatory, and clinical validation work is required before bedside deployment. Challenges include: (1) collecting reproducible breath samples from neonates; (2) managing the highly humid and variable matrix of real exhaled breath; (3) achieving >95% diagnostic accuracy required for screening; (4) device miniaturisation, sterilisation, and single-use or reusable design; (5) integration with clinical workflows in neonatal intensive care or postnatal wards. In Australia, neonatal jaundice is managed according to RANZCOG and RACP/RACP-endorsed guidelines, with serum total bilirubin and transcutaneous bilirubinometry (TcB) as the established diagnostic standards. Exhaled CO breath testing is not currently listed on the Medicare Benefits Schedule (MBS) or recommended in RACGP or RACP clinical guidelines for neonatal jaundice. The TGA would require Class IIb or Class III medical device registration for a diagnostic breath analyser intended for neonatal use, necessitating substantial clinical evidence well beyond this proof-of-concept study. No PBS or MBS pathway exists for this technology at present. Australian neonatal units, particularly in rural and remote settings, could theoretically benefit from a noninvasive, low-cost screening tool if clinical validation is achieved, given the logistical challenges of laboratory bilirubin testing in those environments. Theoretically applicable to neonates with suspected hyperbilirubinaemia where noninvasive CO breath testing could supplement or replace invasive bilirubin sampling; also potentially relevant to adult CO monitoring in occupational or environmental health contexts. Currently applicable only as a research tool.

Abstract

Noninvasive, precise breath analysis holds significant importance for the screening and monitoring of neonatal jaundice, with its core challenge lying in the specific detection of the key biomarker carbon monoxide (CO) within complex breath matrices. This research employed a controlled process to fabricate five sets of In2O3-CuO sensors derived from bimetallic metal-organic frameworks (Bi MOFs). These were assembled into a microsensor array, and by incorporating a linear discriminant analysis algorithm, an electronic nose system was constructed. This approach utilizes thermodynamic feature engineering to enhance data validity and optimize algorithm selection, thereby reducing reliance on large-scale data and computational resources. By integrating thermodynamically feature-driven machine learning, the electronic nose system-comprising merely five In2O3-CuO sensors-ultimately achieved accurate discrimination between CO and acetone. Concurrently, the single 3In2O3-CuO sensor exhibits excellent reproducibility, moisture resistance, and stability, with a low detection limit of 1 ppm and relatively rapid response/recovery times (5/21 s). By introducing machine learning algorithms to analyze the multidimensional response signals from the sensor array, the study successfully addressed the cross-sensitivity issue between CO and coexisting interfering gas C3H6O in clinical environments, achieving qualitative identification of CO ranging from 1 to 50 ppm (cross-validation accuracy reached 82%), providing a highly reliable technical platform for noninvasive screening of neonatal bilirubin metabolic abnormalities.

References

  1. 1.Zhao, F., Jiang, Z., Jiang, X., Wang, R., Yu, L., & Wang, H. (2026). Machine learning-enabled gas sensor based on MOF-derived In2O3-CuO for exhaled CO detection. ACS Sensors. Advance online publication. https://doi.org/10.1021/acssensors.6c00346
Share:XLinkedIn

This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service