Feasibility of nonspinal bony landmark-based image registration in CyberKnife ® robotic radiotherapy for pelvic and femoral bone tumors

Journal of radiation researchMiyabe, Yasuhide, Yamanaka, Hiromu, Seki, Hiroto et al.26 May 2026DOI

Clinical Snapshot

75CEBM
Evidence: Moderateobservational

PICO Framework

P — PopulationPatients with pelvic and femoral bone tumors receiving CyberKnife stereotactic body radiotherapy
I — InterventionNonspinal bony landmark-based image registration (pubic bone and femoral neck as reference points)
C — ComparatorConventional spine-based alignment for image registration
O — OutcomesIntrafractional positional deviations in six degrees of freedom, frequency of intrafractional corrections, alignment accuracy

Bottom Line

This feasibility study suggests that using nonspinal bony landmarks (pubic bone, femoral neck) for image registration in CyberKnife radiotherapy achieves comparable accuracy to conventional spine-based alignment for pelvic and femoral bone tumors. The technique maintained positional accuracy within clinically acceptable limits (±0.5 mm, ±0.5°) with similar correction frequencies. This approach may be particularly valuable when treating tumors anatomically distant from the spine, potentially improving targeting precision. However, the study lacks detailed methodology, statistical analysis, and long-term outcome data. While promising for technical implementation, clinicians should await full publication with comprehensive methodology and larger patient cohorts before routine adoption. The technique appears technically sound but requires validation in diverse patient populations and correlation with clinical outcomes.

Evidence: Moderate

Key Findings

  • P Value: Not reported

  • Effect Size: Non-inferiority demonstrated between nonspinal and spine-based registration

  • Primary Outcome: Intrafractional positional deviations within ±0.5 mm translational and ±0.5° rotational

  • Nnt Or Sensitivity: Clinical accuracy threshold of ±0.5 mm and ±0.5° achieved

  • Confidence Interval: Not reported

Clinical Application

High - technique appears readily implementable in existing CyberKnife centers Relevant to Australian radiation oncology centers with CyberKnife systems; may improve treatment precision for pelvic bone metastases common in prostate and breast cancer patients Patients with pelvic and femoral bone metastases requiring stereotactic radiotherapy

Abstract

In stereotactic body radiotherapy for bone metastases using the CyberKnife system, spinal anatomy is generally employed as the alignment reference. However, when treating pelvic tumors such as those in the pubic bone or proximal femur, the spinal reference may be anatomically distant from the target, potentially compromising visualization and localization accuracy. This study evaluated the feasibility of using nonspinal bony landmarks, specifically the pubic bone and femoral neck, as reference points for image registration. Intrafractional positional deviations in six degrees of freedom and the frequency of intrafractional corrections were analyzed according to treatment site, including cases treated with conventional spine-based alignment. Alignment accuracy was visually and numerically verified by the treatment operator before beam delivery. Across all treatment sites, the mean translational and rotational deviations during treatment remained within ±0.5 mm and ± 0.5°, respectively. Furthermore, no substantial differences were observed in the mean frequency of intrafractional corrections among the different alignment approaches. These findings demonstrate that nonspinal tumor-adjacent bony landmarks can provide clinically acceptable accuracy comparable to conventional spine-based registration. To our knowledge, this is the first study to systematically assess the clinical feasibility of this approach in CyberKnife-based pelvic radiotherapy. The proposed method may serve as a practical and reliable alternative to spine-based alignment, particularly when the spine is remote from the target.

References

  1. 1.Miyabe, Y., Yamanaka, H., Seki, H., Muroi, Y., Nakano, S., Yamazaki, R., Awaya, A., & Mizoe, J. (2026). Feasibility of nonspinal bony landmark-based image registration in CyberKnife robotic radiotherapy for pelvic and femoral bone tumors. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrag028
Share:XLinkedIn

This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service