Fast occupational upper-limb radiation dose prediction using machine learning and Monte Carlo simulation
Clinical Snapshot
PICO Framework
| P — Population | Healthcare professionals performing interventional procedures exposed to scattered radiation |
| I — Intervention | Machine learning ensemble models for radiation dose prediction |
| C — Comparator | Traditional Monte Carlo simulation methods and existing extremity dosimeters |
| O — Outcomes | Accuracy of upper-limb radiation dose prediction (mean absolute error, relative error), computational speed, spatial dose distribution patterns |
Bottom Line
This methodological study demonstrates that machine learning ensemble models can accurately predict upper-limb radiation doses for healthcare workers during interventional procedures, achieving 3.69% relative error compared to Monte Carlo simulations. The rapid prediction capability (10ms) enables real-time dose estimation, potentially revolutionising occupational radiation protection in interventional suites. While the model shows excellent performance under standard conditions, accuracy decreases in extreme beam geometries. The approach addresses limitations of current extremity dosimeters by providing spatial dose information and real-time feedback. Clinical implementation could significantly enhance radiation safety protocols, though validation against real-world dosimeter measurements remains necessary before widespread adoption.
Key Findings
P Value: Not applicable for validation study
Effect Size: 3.69% average relative error
Primary Outcome: Machine learning ensemble model accuracy for upper-limb radiation dose prediction
Nnt Or Sensitivity: Computational speed: ~10 milliseconds per prediction vs hours for full Monte Carlo simulation
Confidence Interval: Not explicitly reported
Clinical Application
High feasibility with existing computational infrastructure; requires integration with procedure planning systems Relevant to Australian radiation protection standards (ACPSEM, ARPANSA guidelines); could enhance compliance with occupational dose monitoring requirements Interventional radiologists, cardiologists, and other healthcare professionals performing image-guided procedures
Abstract
Interventional procedures expose physicians to scattered radiation, particularly to their upper extremities, posing occupational health risks. Existing extremity dosimeters such as thermoluminescents, optically stimulated luminescence rings and active personal dosimeters provide limited spatial information, exhibit angular and energy dependence and offer little or no real-time feedback. This study develops machine learning (ML) models to estimate radiation dose values at discrete upper-limb locations using Monte Carlo (MC)-derived data and procedure-specific parameters. A dataset of 10 000 MC dose maps was generated under varied clinical and geometric conditions. After log-transformation and normalisation, several ML models, including deep neural networks and tree-based regressor, were trained and assessed using five-fold cross-validation. Mean absolute error and relative error (RE) were evaluated on the original dose scale, and an ensemble of the three best-performing models was constructed to improve robustness. The ensemble consistently outperformed individual models, achieving an average RE of 3.69% and demonstrating stable performance across anatomical regions and dose levels. Highest accuracy was obtained for standard beam geometries, whereas larger discrepancies occurred in extreme configurations with steep dose gradients. Predicted dose patterns were consistent with the expected distributions across the upper-limb regions. The findings demonstrate the feasibility of ML-based extremity dose estimation in interventional environments. The proposed ensemble provides a rapid (around 10 ms), scalable alternative to full MC simulations, enabling near real-time predictions of upper-limb occupational dose and supporting optimisation of radiation protection practices in image-guided procedures.
References
- 1.Harb, H., Taguelmimt, K., Benoit, D., Pham, C. H., Nasr, B., & Bert, J. (2026). Fast occupational upper-limb radiation dose prediction using machine learning and Monte Carlo simulation. Journal of Radiological Protection, Article ae63da. https://doi.org/10.1088/1361-6498/ae63da
Related Research
BMJ open
Current evidence on radiofrequency denervation for chronic low back pain: a systematic review and meta-analysis
18 July 2026
Proceedings of the National Academy of Sciences of the United States of America
Magnetically actuated microrobotic system for sequential treatment of biofilm
15 July 2026
ACS nano
Design and Applications of Magnetically Actuated Microrobots in Endovascular Embolization
8 July 2026
This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service