End-to-end testing of an online adaptive radiotherapy workflow for liver tumors using a dynamic Thoraco-abdominal anthropomorphic phantom.

Physics in medicine and biologyBakhtiari Moghaddam, Anahita, Figueiredo Augusto, R, Darremont, P et al.7 Apr 2026DOI

Clinical Snapshot

100CEBM
Evidence: Strongother

PICO Framework

P — PopulationAnthropomorphic phantom simulating liver tumor patients requiring adaptive radiotherapy
I — InterventionTAM-ARa dynamic phantom for end-to-end testing of online adaptive IMRT workflow
C — ComparatorPlanned dose distributions versus measured dose distributions
O — OutcomesDosimetric accuracy, geometric precision, and workflow validation for adaptive radiotherapy

Bottom Line

This validation study demonstrates that the TAM-ARa dynamic anthropomorphic phantom provides a robust platform for end-to-end testing of adaptive radiotherapy workflows for liver tumors. The phantom achieved excellent dosimetric accuracy (<3% deviation) and sub-millimeter geometric precision when validating online adaptive IMRT. The modular design allows systematic testing of various anatomical deformations while maintaining realistic tissue properties across multiple imaging modalities. For radiation oncology centers implementing adaptive radiotherapy, this phantom offers a valuable quality assurance tool that can validate workflow accuracy before clinical implementation. The technology-agnostic design principles could inform development of similar validation tools for other adaptive radiotherapy systems, though this specific validation was limited to the Varian Ethos platform.

Evidence: Strong

Key Findings

  • P Value: Not applicable for validation study

  • Effect Size: Dose deviations reduced to below 3% with online adaptation

  • Primary Outcome: Dosimetric accuracy of adaptive radiotherapy workflow

  • Nnt Or Sensitivity: Sub-millimeter geometric accuracy achieved with full planning target volume coverage

  • Confidence Interval: Not reported for phantom study

Clinical Application

High feasibility for quality assurance programs with modular phantom design allowing rapid setup Relevant for Australian radiation oncology centers adopting adaptive radiotherapy technology, supporting TGA requirements for medical device validation and ACPSEM quality assurance guidelines Radiation oncology centers implementing adaptive radiotherapy for liver tumors

Abstract

Objective.Adaptive radiotherapy (ART) requires robust end-to-end (E2E) testing tools capable of reproducing organ deformation, physiological motion, and multimodal imaging properties. TheToraco-abdominalAnthropomorphic Phantom withMotion forAdaptiveRadiotherapy (TAM-ARa) is a dynamic anthropomorphic phantom developed to support the design, validation, and quality assurance of online ART workflows.Approach.The phantom was constructed with anatomically realistic bone, lung, and abdominal organ models fabricated from tissue equivalent materials. Modular components allowed simulation of interfractional and intrafractional anatomical variations, including ventilator driven respiratory motion, abdominal deformation, and variable gastric filling. CT, CBCT, and 3T MRI scans were acquired to assess imaging performance. Two E2E tests of an online adaptive intensity-modulated radiotherapy (IMRT) workflow for liver tumors were performed on a Varian Ethos (ETHOS, Varian, USA) system. Multiple deformation scenarios were investigated, including a static reference configuration, two abdominal and one thoracic deformation for ionization chamber (IC) dosimetry, and a static plus abdominal deformation scenarios for radiochromic film dosimetry.Main results.The TAM-ARa phantom demonstrated realistic radiological characteristics, with CT Hounsfield units and MRI relaxation times closely matching reportedin vivovalues. The modular architecture allowed reproducible and controlled abdominal deformations, while setup and dosimeter placement were completed within minutes. In E2E tests for adaptive IMRT of liver tumors, excellent agreement was achieved between measured and planned dose distributions after the treatment plan was adapted. IC E2E measurements showed that online adaptation consistently restored accurate dose delivery under deformation, reducing deviations to below 3%. Independent film measurements confirmed sub-millimeter geometric accuracy and full planning target volume coverage after adaptation, demonstrating effective compensation of deformation-induced errors.Significance.The TAM-ARa phantom provides a versatile and reproducible platform for multimodal E2E testing and validation of online ART workflows. Employing its realistic anatomical design, modular structure, and motion capabilities, TAM-ARa was successfully applied for E2E tests to validate a workflow for adaptive IMRT of liver tumors.

References

  1. 1.Bakhtiari Moghaddam, A., Figueiredo Augusto, R., Darremont, P., Runz, A., Echner, G., Johnen, W., Haering, P., Lang, C., Lifferth, M., Qubala, A., Jäkel, O., & Karger, C. P. (2026). End-to-end testing of an online adaptive radiotherapy workflow for liver tumors using a dynamic Thoraco-abdominal anthropomorphic phantom. Physics in Medicine and Biology. https://doi.org/10.1088/1361-6560/ae54f8
Share:XLinkedIn

This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service