Development and validation of a novel deep learning coronary artery plaque quantification model

BMC medical imagingMalmqvist, Johan, Jernberg, Tomas, Vedad, Ramtin et al.14 May 2026DOI

Clinical Snapshot

55CEBM
Evidence: Moderatediagnostic

PICO Framework

P — Population115 patients undergoing coronary computed tomography angiography (CCTA), mean age 58 years, 51% female, mean calcium score 56 Agatston units
I — InterventionQuantiPlaque deep learning software for automated coronary plaque segmentation
C — ComparatorExpert manual annotation as reference standard
O — OutcomesCorrelation and agreement for total and calcified coronary plaque volume quantification using ICC, Pearson's r, Spearman's rho, and Bland-Altman analysis

Bottom Line

This validation study demonstrates that QuantiPlaque deep learning software achieves strong correlation with expert manual annotation for coronary plaque quantification on CCTA (ICC 0.95 for total plaque). However, the study has significant limitations including small sample size (n=115), narrow patient demographics, and lack of external validation. While the correlation metrics are impressive, the clinical utility remains unproven as the study doesn't demonstrate improved patient outcomes or workflow efficiency. The weaker performance in left circumflex territory raises concerns about vessel-specific accuracy. Before clinical implementation, larger multicentre validation studies with diverse patient populations and assessment of clinical impact are needed. Australian clinicians should await TGA approval and local validation before considering adoption.

Evidence: Moderate

Key Findings

  • P Value: Not reported

  • Effect Size: ICC 0.95 for total plaque volume, ICC 0.90 for calcified plaque volume

  • Primary Outcome: Correlation between automated and expert plaque volume quantification

  • Nnt Or Sensitivity: Pearson's r 0.95 for total plaque, 0.90 for calcified plaque; Spearman's rho 0.87 and 0.93 respectively

  • Confidence Interval: 95% limits of agreement: -102 to 85 mm³ for total plaque, -21.97 to 21.42 mm³ for calcified plaque

Clinical Application

Requires integration with existing CCTA workflow and validation in local practice settings Could support CCTA reporting in Australian cardiology and radiology practices, potentially relevant for Medicare item 57360 (CCTA). Would require TGA approval for clinical use and integration with existing PACS systems Patients undergoing CCTA for coronary artery disease assessment, particularly those requiring plaque burden quantification

Abstract

BACKGROUND: Coronary Computed Tomography Angiography (CCTA) is an established tool for assessing coronary artery disease. CCTA determined total coronary artery plaque volume predicts cardiovascular events, but manual quantification is impractical for routine use. Deep learning-based methods offer a promising solution for automated plaque volume assessment. METHODS: We developed and validated a novel software, QuantiPlaque, powered by deep learning models, for automated segmentation of total and calcified coronary plaque volume. Expert manual annotation served as the reference standard. Correlations between deep learning and expert segmentation were evaluated using intraclass correlation coefficient (ICC), Pearson's r, and Spearman's rho, and agreements were evaluated with Bland-Altman analyses. RESULTS: A total of 115 CCTA scans were included. Mean (range) age was 58 (51-64) and 51% were females. The mean coronary artery calcium score was 56 Agatston units, ranging from 0 to 601. The model demonstrated strong correlation and agreement with expert annotation for per-patient total plaque volume (ICC 0.95, mean difference - 8.35 mm3, 95% limit of agreement - 102 to 85 mm3, Spearman's rho 0.87, Pearson's r 0.95) and calcified plaque volume (ICC 0.90, mean difference - 0.28 mm3, 95% limits of agreement - 21.97 to 21.42 mm3, Spearman's rho 0.93, Pearson's r 0.90). Per-vessel analysis showed strong correlation in the left anterior descending artery and right coronary artery but was weaker in the left circumflex artery territory. CONCLUSION: The evaluated deep learning model provides accurate quantification of total and calcified plaque burden, with strong correlation and agreement to expert annotation.

References

  1. 1.Malmqvist, J., Jernberg, T., Vedad, R., & Wang, C. (2026). Development and validation of a novel deep learning coronary artery plaque quantification model. BMC Medical Imaging. https://doi.org/10.1007/s00330-025-11410-w
Share:XLinkedIn

This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service