Research Appraisalobservational

Computer modeling to improve measurement of stroke volume with echocardiography

Cardiovascular ultrasoundMurthi, Sarah Balderston, Yang, Shiming, Oliveri, Peter P et al.6 July 2026DOI

Clinical Snapshot

50CEBM
Evidence: Weakobservational

PICO Framework

P — PopulationCritically ill adult patients with a pulmonary artery catheter (PAC) placed for clinical indications, enrolled prospectively over 20 months at a single academic centre
I — InterventionEchocardiographic stroke volume (SV) calculated using a machine learning computer model estimate of left ventricular outflow tract diameter (LVOTdCM) combined with velocity time integral (VTI)
C — ComparatorTwo comparators: (1) SV calculated using human expert measurement of LVOTd (LVOTdHEM) combined with VTI; (2) SV measured simultaneously by pulmonary artery catheter (PAC) thermodilution as the reference standard
O — OutcomesAgreement between echocardiographic SV estimates (CM and HEM) and PAC-derived SV assessed by Bland-Altman analysis (mean bias and 95% limits of agreement); diagnostic yield (proportion of patients in whom SV could be calculated)

Bottom Line

This prospective single-centre study of 84 critically ill patients with pulmonary artery catheters demonstrates that a machine learning model for estimating left ventricular outflow tract diameter (LVOTd) can increase echocardiographic stroke volume measurement yield from 70% to 92% — a clinically meaningful 22% absolute improvement — while maintaining agreement with PAC thermodilution comparable to human expert measurement. Both methods showed similar mean bias (approximately 1.75–3.1 mL) relative to PAC, but the 95% limits of agreement were wide for both (approximately ±30 mL), which is a clinically important caveat for individual patient management decisions. The study is limited by its small sample, single-centre design, absence of explicit blinding protocols, and lack of clinical outcome data. The machine learning model is not fully characterised in this paper. For Australian intensivists and emergency physicians, the concept of simplifying echocardiographic SV assessment is attractive, particularly as PAC use declines. However, the wide measurement variability observed means clinicians should interpret individual SV estimates with appropriate caution. Larger, multicentre validation studies with clinical outcome endpoints are needed before this approach can be recommended for routine clinical implementation.

Evidence: Weak

Key Findings

  • P Value: Not reported in abstract

  • Effect Size: Mean bias: HEM vs PAC = 3.1 mL; CM vs PAC = 1.75 mL. Diagnostic yield: HEM = 70% (59/84); CM = 92% (78/84) — absolute improvement of 22%

  • Primary Outcome: Agreement between echocardiographic SV (HEM and CM methods) and PAC-derived SV by Bland-Altman analysis, and diagnostic yield (proportion of patients with calculable SV)

  • Nnt Or Sensitivity: Diagnostic yield improvement: NNT-equivalent = approximately 4.5 patients assessed to gain one additional calculable SV measurement with CM vs HEM. No sensitivity/specificity reported (continuous measurement study). Percentage error for HEM vs PAC and CM vs PAC not explicitly calculated in abstract but LOA suggest clinically wide agreement intervals.

  • Confidence Interval: 95% limits of agreement: HEM vs PAC = -28.8 to +34.9 mL (total range 63.7 mL); CM vs PAC = -29.0 to +32.5 mL (total range 61.5 mL). Confidence intervals for mean bias not reported.

Clinical Application

The computer model requires only VTI acquisition (a single Doppler measurement) rather than both LVOTd and VTI, simplifying the workflow and potentially enabling less experienced operators to obtain quantitative SV estimates. However, VTI acquisition itself requires adequate apical five-chamber or three-chamber views and Doppler alignment, which may still be technically challenging in critically ill patients. Full clinical implementation requires integration of the model into echocardiography platforms, which is not yet described as commercially available. Point-of-care echocardiography is increasingly used in Australian ICUs and emergency departments, supported by CICM (College of Intensive Care Medicine) and ACEM (Australasian College for Emergency Medicine) training frameworks. The RACGP does not have specific guidelines for echocardiographic SV measurement in primary care. There is no current TGA-listed or PBS-subsidised machine learning echocardiography tool for LVOTd estimation in Australia. The findings are relevant to Australian intensivists and emergency physicians who perform or interpret point-of-care echocardiography, particularly in resource-limited settings where PAC use is declining and non-invasive haemodynamic assessment is increasingly relied upon. Australian ICU practice increasingly favours less invasive haemodynamic monitoring, making accurate echocardiographic SV measurement clinically important. Adoption would require TGA regulatory approval for any commercial software implementation. Critically ill adult patients in intensive care settings where haemodynamic monitoring is required and echocardiographic SV assessment is clinically indicated. Most directly applicable to patients already receiving invasive haemodynamic monitoring (PAC or equivalent). Potential broader application to perioperative and emergency department settings where point-of-care echocardiography is performed.

Abstract

BACKGROUND: The stroke volume (SV) can be measured by a human expert (HE) using the left ventricular outflow tract diameter (LVOTd) and its velocity time integral (VTI) with echo. If the SV is known, the cardiac output and systemic vascular resistance can be calculated. We have previously described a machine learning computer model (CM) to estimate the LVOTd. The accuracy of LVOTdCM vs. LVOTdHEM in estimation of the SV relative to a pulmonary artery catheter (PAC) is unknown. METHOD: Over 20 months, a prospective observational study was done in patients with a PAC placed for clinical indications. Participants underwent an echocardiogram, and at the same time the SV was measured by PAC. The SV with echo was calculated using the LVOTdHEM and LVOTdCM and the VTI. The bias and agreement between the metrics and PAC were determined with Bland-Altman analysis. RESULTS: Computer modeling improves measurement of stroke volume with echocardiography. Eighty-four patients with a PAC were enrolled. With human expert measurement of the LVOTd and the VTI, the SV could be calculated in 59 (70%), and with the LVOTdCM and the VTI in 78 (92%). Bland Altman analysis comparing PAC and HEM yielded a mean bias of 3.1 with 95% limits of agreement (LOA) -28.8 and 34.9. When comparing PAC and CM, bias was 1.75 and LOA of -29 and 32.5. CONCLUSIONS: A machine learning model of the LVOTd allows accurate calculation of the SV using only the VTI, simplifying the assessment while increasing the yield by 22%. Machine learning tools can lower the bar for obtaining quantitative metrics with echo. Adding objective and repeatable measures will improve the ability of echo to guide therapy in the critically ill.

Share:XLinkedIn

This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service