Bluetooth low energy thermohygrometric sensor for continuous respiratory rate monitoring: 48 h breathing simulator and human proof-of-concept
Clinical Snapshot
PICO Framework
| P — Population | Breathing simulator (48h validation) and three healthy adults |
| I — Intervention | Bluetooth Low Energy thermohygrometric sensor (AHT21 + Nordic nRF52840) attached to oxygen mask |
| C — Comparator | Capnography (gold standard) and programmed breathing simulator rates |
| O — Outcomes | Respiratory rate measurement accuracy, correlation with reference standards, operational limits |
Bottom Line
This proof-of-concept study demonstrates promising technical accuracy for a novel Bluetooth-enabled respiratory rate monitor using temperature-humidity sensing. The device achieved 96.4% accuracy in controlled simulator testing and showed good agreement with capnography in three healthy adults. However, the clinical evidence remains very limited - only three healthy volunteers were tested, with no validation in actual patients or real clinical environments. Key limitations include sensor saturation at high respiratory rates and humidity levels, and reduced performance at low oxygen flow rates. While the technology shows potential for low-cost respiratory monitoring in resource-limited settings, substantial further validation in diverse patient populations and clinical scenarios is essential before clinical implementation. Australian clinicians should await larger validation studies and TGA approval before considering clinical adoption.
Key Findings
P Value: P < 0.0001 for correlation with programmed rates
Effect Size: Strong correlation with reference standards (r = 0.996 simulator, mean bias 0.44 breaths/min vs capnography)
Primary Outcome: 96.4% overall breath detection accuracy in 48h simulator validation
Nnt Or Sensitivity: Sensitivity reduced at high respiratory rates (60 breaths/min) and high ambient humidity (90% RH)
Confidence Interval: 95% limits of agreement: -1.86 to 2.75 breaths/min
Clinical Application
Low-cost, portable device with Bluetooth connectivity offers practical advantages over capnography Developed by Australian institutions (Griffith University, University of Queensland, Wesley Private Hospital). Could address monitoring gaps in rural/remote settings and aged care facilities where capnography unavailable. TGA approval would be required for clinical use. Potentially suitable for stable patients requiring continuous respiratory monitoring in low-acuity settings
Abstract
Continuous respiratory rate monitoring is essential for early detection of clinical deterioration, but conventional methods like capnography are often impractical outside high-acuity settings. We developed a low-cost, Bluetooth Low Energy (BLE) respiratory rate monitor incorporating a MEMS temperature-humidity sensor (AHT21) and Nordic nRF52840 microcontroller, designed to attach externally to a standard oxygen mask. The device detects breaths using rapid, cyclic changes in exhaled temperature and humidity via a slope-based time-domain algorithm. Bench validation over 48 h using a breathing simulator (5-60 breaths min-1, tidal volume 300 ml, 34 °C, 95%-100% RH) demonstrated 96.4% overall breath detection accuracy and strong correlation with programmed rates (r = 0.996, P < 0.0001). Sensor saturation occurred at high respiratory rates (60 breaths min-1), and high ambient humidity levels (90% RH) both of which reduced measurement accuracy. Human testing in three healthy adults showed strong agreement with capnography (mean bias 0.44 breaths min-1; 95% limits of agreement -1.86 to 2.75), with minor under-detection at the highest rate (50 breaths min-1). The device operated reliably across oxygen flow rates of 4-15 l min-1, with reduced performance at 0-2 l min-1due to saturation of the sensor. These results demonstrate that BLE-enabled MEMS thermohygrometric sensors can accurately monitor respiratory rate under controlled conditions while highlighting operational limits due to sensor saturation. This study extends prior work on humidity-based respiratory monitoring and illustrates the potential for low-cost, portable respiratory rate devices suitable for low-acuity or resource-limited clinical settings.
Related Research
Sleep medicine
Concordance of wearable device sleep metrics with patient-reported sleep quality: A systematic review
3 Aug 2026
Bioelectrochemistry (Amsterdam, Netherlands)
From data to diagnosis: A comprehensive review of machine learning-driven wearable sensors in healthcare.
2 Aug 2026
JMIR mHealth and uHealth
Wearable Devices for Monitoring and Management of Comorbid Obstructive Sleep Apnea and Hypertension: Scoping Review
1 Aug 2026
This content is for educational purposes for healthcare professionals only and does not constitute clinical advice. Clinical decisions should be based on individual patient assessment, current guidelines, and appropriate specialist consultation. Editorial Standards · Privacy Policy · Terms of Service